On arXiv: Reducing the hydrogen content in liquid helium
Molecular hydrogen can block liquid He (LHe) intake capillaries typically used on 1K pots of many cryostats, causing severe disruption to the experiments. Here, we measure molecular hydrogen content in the liquifier helium gas supply and present measures to decrease hydrogen content in LHe: (i) better He supply quality (ii) release of dirt gas during liquifier purge cycles and (iii) installation of two catalytic converters to absorb the hydrogen. These actions have eliminated our low-temperature impedance blockage occurrences now for more than two years.
Sifrig et al., arXiv:2010:14136 (Oct 27, 2020)